DocumentCode
3787138
Title
The contribution of blood-flow-induced conductivity changes to measured impedance
Author
J. Wtorek;A. Polinski
Author_Institution
Tech. Univ. Gdansk, Poland
Volume
52
Issue
1
fYear
2005
Firstpage
41
Lastpage
49
Abstract
The paper considers the contribution of conductivity changes undergone in an anisotropical medium to measured resistance. This was achieved by extending the relationship proposed by Geselowitz to anisotropical materials described, therefore, by a conductivity tensor. It was found that each element of a conductivity change tensor contributed to the measured resistance only if a corresponding component of the electrical field was nonzero. Numerical calculations were performed for blood-flow-associated conductivity changes. A special experiment stand was developed which allowed experiments to be performed proving the theoretical results. It was found that the absolute value of resistance change measured in the direction perpendicular to the vessel axis was much smaller than that measured along the vessel axis. The results obtained may explain the fact that the actual change of measured resistance created by changes of conductivity induced by aortic blood flow is lower than expected from simplified models.
Keywords
"Conductivity measurement","Impedance measurement","Electrical resistance measurement","Electric resistance","Anisotropic magnetoresistance","Tensile stress","Conducting materials","Electric variables measurement","Fluid flow measurement","Blood flow"
Journal_Title
IEEE Transactions on Biomedical Engineering
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2004.839633
Filename
1369587
Link To Document