• DocumentCode
    380471
  • Title

    Non-invasive measurement of diaphragmatic contraction timing in dogs

  • Author

    Torres, A. ; Fiz, J.A. ; Morera, J. ; Grassino, A.E. ; Jané, R.

  • Author_Institution
    Centre de Recerca en Enginyeria Biomedica, Univ. Politecnica de Catalunya, Barcelona, Spain
  • Volume
    1
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    437
  • Abstract
    The movement of thoracic cage (TM) measured with a piezoelectric contact sensor placed on the costal wall is presented in this work as a new non-invasive technique for diaphragmatic contraction period (CP) monitoring. The relationship between CP estimated with this technique is compared with estimations done with other respiratory signals commonly employed for physiological research studies: diaphragm length measured with sonomicrometry (DL), transdiaphragmatic pressure (DP) and respiratory airflow (FL). Specific algorithms were developed to determinate the CP in the four signals. Experiments were performed in three pentotarbital-anesthetized mongrel dogs. Two respiratory tests were studied: spontaneous ventilations (SV) and respiration with an inspiratory load (IL). CP estimated with DL signal was used as reference because this signal is the directly related with the diaphragmatic contraction. Different parameters were estimated for the study of the relationship between CP measured with DL signal and CP measured with TM, FL and DP signals. High relationships were obtained in the IL respiratory test. Lower values were obtained in SV protocol, but the parameters obtained from TM signal correlated better with the ones obtained from DL signal than those from FL and DP signals. Results confirm that it is possible to monitor the diaphragmatic contraction timing with a non-invasive piezoelectric contact sensor.
  • Keywords
    biomechanics; biomedical measurement; mechanical variables measurement; patient monitoring; piezoelectric transducers; pneumodynamics; diaphragm; diaphragmatic contraction timing; dogs; inspiratory load; noninvasive measurement; pentotarbital-anesthetized mongrel dogs; piezoelectric contact sensor; respiratory airflow; respiratory monitoring; signal parameters; sonomicrometry; thoracic cage movement; transdiaphragmatic pressure; Biomedical monitoring; Biosensors; Change detection algorithms; Condition monitoring; Dogs; Frequency; Hospitals; Length measurement; Testing; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-7211-5
  • Type

    conf

  • DOI
    10.1109/IEMBS.2001.1018956
  • Filename
    1018956