• DocumentCode
    381796
  • Title

    A novel fluidic network system for enzyme batch-immobilization and blood sensing

  • Author

    Tseng, Fan-Gang ; Lin, Kuang-Hua ; Chieng, Ching-Chang

  • Author_Institution
    ESS Dept., Nat. Tsing Hua Univ., Hsinchu, Taiwan
  • Volume
    1
  • fYear
    2002
  • fDate
    2002
  • Firstpage
    689
  • Abstract
    This paper reports a novel fluidic network system for immobilizing various enzymes on arrays in parallel for blood sensing. The network was fabricated by polymer-MEMS technology with plasma treatment on channel surface, and the liquids were driven by surface tension passively without external active power source. Enzyme based glucose sensing was employed as a model system in this fluidic network, and the testing result showed precise enzyme immobilization and functional sensor response. The fully fabricated glucose sensor in this fluidic network system demonstrated a sensitivity of 5.42 nA/mM, and a response linearly related to the glucose concentration. This novel fluidic network can be applied to blood sensing for clinical or domestic usage.
  • Keywords
    biosensors; blood; microfluidics; organic compounds; proteins; surface tension; blood sensing; clinical usage; domestic usage; enzyme batch-immobilization; external active power source; fluidic network system; functional sensor response; glucose concentration; linearly related response; plasma treatment; polymer-MEMS technology; precise enzyme immobilization; surface tension; Biochemistry; Blood; Liquids; Plasma sources; Polymers; Power system modeling; Sensor systems; Sugar; Surface tension; Surface treatment;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2002. Proceedings of IEEE
  • Print_ISBN
    0-7803-7454-1
  • Type

    conf

  • DOI
    10.1109/ICSENS.2002.1037186
  • Filename
    1037186