    The performance of multi pixel photon counters (MPPC) of 3 mm × 3 mm size, with 14400 and 3600 pixels, were studied by means of the signal from a laser light pulser and using the 3 mm × 3 mm × 20 mm LSO pixel scintillator. Special attention was paid to measure number of fired pixels, generated by the light of pulser and that of the LSO crystal, using a direct method of a comparison of the light peak position in the pulse height spectrum with that of the single photoelectron. The tests of the LSO crystal showed 1550 ± 80 fired pixels per MeV in the MPPC with 14400 pixels assuring a good linearity of the response up to about 1 MeV energy of gamma rays absorbed in the LSO crystal. Energy resolution of 14.8% for 662 keV gamma rays from 137Cs source and a time resolution of about 850 ps for 511 keV annihilation quanta were limited by a rather low number of the fired pixels compared to the number of photoelectrons in photomultipliers.
  • Keywords
    "Counting circuits","Positron emission tomography","Pulse measurements","Radiation detectors","Gamma rays","Energy resolution","Optical pulses","Position measurement","Optical pulse generation","Testing"
  • Journal_Title
    IEEE Transactions on Nuclear Science
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2010.2044586
  • Filename
    5485129