DocumentCode :
3853739
Title :
Nondestructive testing ultrasonic immersion probe assessment and uncertainty evaluation according to EN 12668-2:2010
Author :
C. E. R. Silva;A. V. Alvarenga;R. P. B. Costa-Felix
Author_Institution :
Lab. of Ultrasound, Nat. Inst. of Metrol., Quality &
Volume :
59
Issue :
10
fYear :
2012
Abstract :
Ultrasonic beam parameters from nondestructive testing probes and respective measurement uncertainties were calculated according to the Guide to the Expression of Uncertainty in Measurement (BIPM-JCGM-100:2008). The major conclusion of this work is that proper measurements of ultrasonic probes parameters are necessary because each probe has intrinsic construction particularities. Uncertainty evaluation was essential to properly assess the experimental results.
Keywords :
"Probes","Uncertainty","Acoustics","Ultrasonic variables measurement","Measurement uncertainty","Structural beams","Attenuation"
Journal_Title :
IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2012.2459