• DocumentCode
    385410
  • Title

    Configuration of the mitral valves subvalvular complex and its effect on the chordal force distribution

  • Author

    Soerensen, D.D. ; Christensen, T.B.N. ; He, Z. ; He, S. ; Yoganathan, A.P.

  • Author_Institution
    Wallace H. Coulter Dept. of Biomed. Eng., Georgia Inst. of Technol., Atlanta, GA, USA
  • Volume
    2
  • fYear
    2002
  • fDate
    2002
  • Firstpage
    1246
  • Abstract
    Three mitral valves (MV) with annulus, leaflets, chordae tendineae (CT) and papillary muscles (PM) were extracted intact from fresh porcine hearts. The valves were inserted in an in vitro left heart simulator that provided physiological flow and transmitral pressures. Four C-shaped chordal force transducers were attached to four chordae tendineae; anterior strut, basal posterior, marginal posterior stem and a commissure. All four chordae originated from the posteromedial papillary muscle. The force transducers measured the force exerted on the four individual chordae during cardiac cycles, under different conditions: two peak transmitral pressures (120 mmHg and 150 mmHg) and three papillary muscle positions (normal, taut and slack). Taut was 5 mm from normal position, and slack 3 mm from normal position. The chordal force distribution changed with papillary muscle displacement. The anterior strut and the basal posterior chordae bore the biggest tension in normal and taut position, whereas the posterior marginal stem bore most of the tension in the slack papillary muscle position. Increasing transmitral pressure increased the magnitude of the chordal forces, but not the force distribution between the chordae.
  • Keywords
    biomechanics; biomedical transducers; cardiology; force sensors; muscle; 120 mmHg; 150 mmHg; C-shaped chordal force transducers; annulus; anterior strut; basal posterior stem; cardiac cycles; chordae tendineae; chordal force distribution; commissure; force distribution; force transducers; fresh porcine hearts; in vitro left heart simulator; leaflets; marginal posterior stem; mitral valves; normal position; papillary muscle displacement; papillary muscle positions; papillary muscles; peak transmitral pressures; physiological flow; posteromedial papillary muscle; slack position; subvalvular complex configuration; taut position; transmitral pressures; Biomedical measurements; Boring; Capacitive sensors; Displacement measurement; Force measurement; Heart valves; Helium; Muscles; Position measurement; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology, 2002. 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society EMBS/BMES Conference, 2002. Proceedings of the Second Joint
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-7612-9
  • Type

    conf

  • DOI
    10.1109/IEMBS.2002.1106370
  • Filename