DocumentCode :
3861091
Title :
A novel vector network analyzer
Author :
K. Hoffmann;Z. Skvor
Author_Institution :
Fac. of Electr. Eng., Czech Tech. Univ., Prague, Czech Republic
Volume :
46
Issue :
12
fYear :
1998
Firstpage :
2520
Lastpage :
2523
Abstract :
A calibration procedure for a perturbation two-port vector network analyzer is presented. It consists of a variable perturbation two-port placed between a device-under-test and a scalar network analyzer. A measured vector reflection coefficient is determined on the basis of amplitude-only measurements. A full correction of systematic errors is possible. The new principle was experimentally confirmed in the frequency band up to 14 GHz. The results enable prediction of an add-on for scalar analyzers, enabling vector measurements.
Keywords :
"Calibration","Error correction","Reflection","Measurement errors","Frequency","Measurement standards","Wideband","Costs","Information analysis","Bridges"
Journal_Title :
IEEE Transactions on Microwave Theory and Techniques
Publisher :
ieee
ISSN :
0018-9480
Type :
jour
DOI :
10.1109/22.739243