• DocumentCode
    3861791
  • Title

    Design of matrix shim coils system for nuclear magnetic resonance

  • Author

    P. Konzbul;K. Sveda;A. Srnka

  • Author_Institution
    Inst. of Sci. Instrum., Czechoslovak Acad. of Sci., Brno, Czech Republic
  • Volume
    36
  • Issue
    4
  • fYear
    2000
  • Firstpage
    1732
  • Lastpage
    1735
  • Abstract
    A high purity magnetic field matrix shim system for Nuclear Magnetic Resonance has been designed and constructed. The stochastic methods, suitably determined by the parameters obtained from the theoretical analysis, are used for optimization of its coils. The approach to the design ensures the high purity of correcting magnetic fields and minimizes the total shim system power consumption. The measuring technique for the magnetic field space configuration of the matrix shim system for NMR is also presented. The used magnetic induction coil method with the optimal measuring procedure and data analysis are suitable for checking the manufacturing technology.
  • Keywords
    "Coils","Nuclear magnetic resonance","Magnetic field measurement","Space technology","Magnetic fields","Stochastic processes","Magnetic analysis","Optimization methods","Energy consumption","Data analysis"
  • Journal_Title
    IEEE Transactions on Magnetics
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.877778
  • Filename
    877778