• DocumentCode
    3862797
  • Title

    Nanophotonic lab-on-a-chip Raman sensors: A sensitivity comparison with confocal Raman microscope

  • Author

    Ashim Dhakal;Pieter Wuytens;Frederic Peyskens;Ananth Subramanian;Andre Skirtach;Nicolas Le Thomas;Roel Baets

  • Author_Institution
    Photonics Research Group, Department of Information Technology, Ghent University-IMEC, Ghent (Belgium)
  • fYear
    2015
  • fDate
    5/15/2016 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    We compare the performance of the nano-photonic waveguide based evanescent Raman sensors and a typical the confocal microscope. For 1 cm long strip silicon nitride waveguides, we theoretically expect more than 500 times higher signal compared to the confocal microscopic systems. The results of our preliminary measurements indicate at least 50 times higher signal. For slotted waveguides, the preliminary measurement result yields about 300 times higher signal compared to confocal microscope while the theoretical figure is more than 2000.
  • Keywords
    "Optical waveguides","Microscopy","Waveguide theory","Raman scattering","Strips"
  • Publisher
    ieee
  • Conference_Titel
    BioPhotonics (BioPhotonics), 2015 International Conference on
  • Type

    conf

  • DOI
    10.1109/BioPhotonics.2015.7304021
  • Filename
    7304021