DocumentCode :
393653
Title :
Infrared radiation thermometry near room temperature at NMIJ
Author :
Ishii, J. ; Kaneko, Y. ; Shimizu, Y. ; Sakuma, F.
Volume :
4
fYear :
2002
fDate :
5-7 Aug. 2002
Firstpage :
2070
Abstract :
Facilities and calibration techniques for infrared radiation thermometry in the low temperature range have been developed at the NMIJ. The estimated uncertainties (k = 2) of the radiance temperature of the reference blackbody sources are better than 60 mK in the temperature range from -20°C to 160°C. A compact infrared radiation thermometer has been developed as a transfer instrument. The uncertainty of calibration of the industrial standard blackbodies has been estimated to be smaller than 0.1°C.
Keywords :
blackbody radiation; calibration; measurement uncertainty; spectral methods of temperature measurement; transfer standards; -20 to 160 degC; calibration; industrial reference blackbody source; infrared radiation thermometry; measurement uncertainty; room temperature; transfer instrument; Calibration; Coils; Copper; Inspection; Nitrogen; Optical sensors; Ray tracing; Temperature distribution; Temperature sensors; Uncertainty;