• DocumentCode
    393653
  • Title

    Infrared radiation thermometry near room temperature at NMIJ

  • Author

    Ishii, J. ; Kaneko, Y. ; Shimizu, Y. ; Sakuma, F.

  • Volume
    4
  • fYear
    2002
  • fDate
    5-7 Aug. 2002
  • Firstpage
    2070
  • Abstract
    Facilities and calibration techniques for infrared radiation thermometry in the low temperature range have been developed at the NMIJ. The estimated uncertainties (k = 2) of the radiance temperature of the reference blackbody sources are better than 60 mK in the temperature range from -20°C to 160°C. A compact infrared radiation thermometer has been developed as a transfer instrument. The uncertainty of calibration of the industrial standard blackbodies has been estimated to be smaller than 0.1°C.
  • Keywords
    blackbody radiation; calibration; measurement uncertainty; spectral methods of temperature measurement; transfer standards; -20 to 160 degC; calibration; industrial reference blackbody source; infrared radiation thermometry; measurement uncertainty; room temperature; transfer instrument; Calibration; Coils; Copper; Inspection; Nitrogen; Optical sensors; Ray tracing; Temperature distribution; Temperature sensors; Uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SICE 2002. Proceedings of the 41st SICE Annual Conference
  • Print_ISBN
    0-7803-7631-5
  • Type

    conf

  • DOI
    10.1109/SICE.2002.1195713
  • Filename
    1195713