DocumentCode :
39733
Title :
NPLI Cesium Atomic Fountain Frequency Standard: Preliminary Results
Author :
Arora, Pooja ; Purnapatra, S.B. ; Acharya, Arup ; Agarwal, Abhishek ; Yadav, Suneel ; Pant, Kapil ; Gupta, Ananya Sen
Author_Institution :
Time & Freq. Sect., Council of Sci. & Ind. Res.-Nat. Phys. Lab., New Delhi, India
Volume :
62
Issue :
7
fYear :
2013
fDate :
Jul-13
Firstpage :
2037
Lastpage :
2042
Abstract :
National Physical Laboratory India has developed a cesium atomic fountain frequency standard. The fountain has now become operational and is ready for complete frequency evaluation. In this paper, we report for the first time, the preliminary results obtained with this fountain, which include Ramsey fringes, frequency stability analysis, with an H-maser, and magnetic field mapping in the atomic flight region.
Keywords :
atomic clocks; frequency standards; H-maser; NPLI; National Physical Laboratory India; Ramsey fringes; atomic flight region; cesium atomic fountain frequency standard; complete frequency evaluation; frequency stability analysis; magnetic field mapping; Atomic clocks; Ramsey fringes; microwave frequency standard;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2013.2255974
Filename :
6509912
Link To Document :
بازگشت