• DocumentCode
    399070
  • Title

    Precise calibration for optical amplifier noise figure measurement using the RIN subtraction method

  • Author

    Obarski, G.

  • Author_Institution
    NIST, Boulder, CO, USA
  • fYear
    2003
  • fDate
    23-28 March 2003
  • Firstpage
    601
  • Abstract
    Calibration with a relative-intensity noise (RIN) standard significantly improves the RIN subtraction method for measuring the noise figure of an erbium-doped fiber amplifier (EDFA). The typical standard uncertainty in our measurement is ≤0.15 dB.
  • Keywords
    erbium; noise measurement; optical fibre amplifiers; optical loss measurement; wavelength division multiplexing; RIN subtraction method; erbium-doped optical amplifier; noise figure measurement; precise calibration; relative-intensity noise; Calibration; Erbium-doped fiber amplifier; Measurement standards; Measurement uncertainty; Noise figure; Noise measurement; Optical amplifiers; Optical noise; Semiconductor optical amplifiers; Stimulated emission;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Optical Fiber Communications Conference, 2003. OFC 2003
  • Print_ISBN
    1-55752-746-6
  • Type

    conf

  • DOI
    10.1109/OFC.2003.1248441
  • Filename
    1248441