• DocumentCode
    410162
  • Title

    Zone-based B-mode imaging

  • Author

    Napolitano, David ; Chou, Ching-Hua ; McLaughlin, Glen W. ; DeBusschere, Derek ; Mo, Larry Y L ; Ji, Ting-Lan

  • Author_Institution
    ZONARE Med. Syst. Inc., Mountain View, CA, USA
  • Volume
    1
  • fYear
    2003
  • fDate
    5-8 Oct. 2003
  • Firstpage
    25
  • Abstract
    A zone-based technique for real-time B-mode imaging is described. The technique utilizes a broad transmit beam from which many receive beams are formed, such that a full field of view image can be formed using only 5-15 firings. On receive, the RF data is pre-processed and accumulated in a channel domain baseband I/Q memory, and then transferred to a DSP-based imaging system, which performs dynamic receive focusing, detection, log compression, spatial filtering, and scan conversion. This technique and architecture extracts more information from each transmit firing, transforming the image formation rate problem from one of acoustic propagation time limitations, to processing speed limitations and thus, leverages Moore´s Law. The basic technique and architecture will be discussed, as well as providing several example images from different applications.
  • Keywords
    acoustic focusing; acoustic signal processing; ultrasonic imaging; DSP based imaging system; Moores law; acoustic propagation; broad transmit beam; channel domain baseband I/Q memory; image formation rate; log compression; real time B-mode imaging; scan conversion; spatial filtering; speed limitations; zone based B-mode imaging; Acoustic beams; Acoustic imaging; Acoustic signal detection; Baseband; Data mining; Filtering; Focusing; Image coding; Image converters; Radio frequency;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics, 2003 IEEE Symposium on
  • Print_ISBN
    0-7803-7922-5
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2003.1293349
  • Filename
    1293349