• DocumentCode
    41374
  • Title

    Selectivity for Specific Cardiovascular Effects of Vagal Nerve Stimulation With a Multi-Contact Electrode Cuff

  • Author

    Ordelman, S.C.M.A. ; Kornet, L. ; Cornelussen, R. ; Buschman, H.P.J. ; Veltink, Peter H.

  • Author_Institution
    MIRA Inst. for Biomed. Technol. & Tech. Med., Univ. of Twente, Enschede, Netherlands
  • Volume
    21
  • Issue
    1
  • fYear
    2013
  • fDate
    Jan. 2013
  • Firstpage
    32
  • Lastpage
    36
  • Abstract
    The cardiovascular system can be influenced by electrically stimulating the vagal nerve. Selectivity for specific cardiac fibers may be limited when stimulating at the cervical level. Our objective was to increase effectiveness and selectivity for cardiovascular effects of vagal nerve stimulation by using local bipolar stimulation in one nerve cross section using a multi-contact cuff instead of less localized stimulation using a tripolar ring electrode. Both types of cuff electrodes were compared with respect to their relative effects on R-R interval (RRI), P-Q interval (PQI), left ventricular contractility (LVC), and left ventricular pressure (PLV) in seven pigs. Stimulation using the optimal bipolar configuration on the multi-contact cuff significantly affected RRI, PQI, LVC, and PLV, whereas stimulation with the ring electrode only significantly affected RRI and PQI. The cardiovascular parameters that could be significantly influenced varied between the bipolar configurations. These novel findings may be relevant for optimizing electrode configurations for clinical cardiac applications of vagal nerve stimulation.
  • Keywords
    bioelectric phenomena; biomedical electrodes; cardiovascular system; neurophysiology; LVC; P-Q interval; PQI; R-R interval; RRI; bipolar configuration; cardiac fibers; cardiovascular effects; cardiovascular parameters; cardiovascular system; cervical level; clinical cardiac applications; electrical stimulation; electrode configurations; left ventricular contractility; left ventricular pressure; local bipolar stimulation; multicontact electrode cuff; nerve cross section; ring electrode; vagal nerve stimulation; Cathodes; Contacts; Electrical stimulation; Heart rate; Rail to rail inputs; Biomedical electrodes; cardiovascular system; electrical stimulation; vagus nerve; Animals; Blood Pressure; Electric Stimulation; Equipment Design; Equipment Failure Analysis; Female; Heart Rate; Implantable Neurostimulators; Myocardial Contraction; Sensitivity and Specificity; Swine; Vagus Nerve; Ventricular Function;
  • fLanguage
    English
  • Journal_Title
    Neural Systems and Rehabilitation Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1534-4320
  • Type

    jour

  • DOI
    10.1109/TNSRE.2012.2214058
  • Filename
    6298966