Abstract :
Recent progress in magnetic resonance electrical impedance tomography (MREIT) research has shown that conductivity images with higher spatial resolution and accuracy are achievable. One of the most important remaining problems to be solved in MREIT before we can apply the technique to human subjects is how to reduce the amount of injection current. Since we use an MRI scanner to measure the induced magnetic flux density data subject to an injection current, the data is contaminated with random noise. In order to obtain enough signal-to-noise ratio (SNR), we need to inject a large amount of current into the subject. However, it is obvious that we must comply with the electrical safety regulations and this means that we should deal with noisy data having a low SNR due to the limited amount of injection current. Furthermore, in the developed reconstruction algorithms, the required numerical differentiations of the noisy data may result in deterioration of the reconstructed conductivity image leading to a loss of important information. We propose a PDE-based denoising technique that diminishes the degradation of reconstructed conductivity images due to the noise in measured data. The proposed PDE-based technique is advantageous in reducing the random noise while preserving useful features in MREIT.
Keywords :
biomedical MRI; biomedical equipment; electric impedance imaging; image denoising; image reconstruction; image resolution; image scanners; medical image processing; partial differential equations; random noise; tomography; MRI scanner; PDE; conductivity image reconstruction; denoising technique; electrical safety regulation; injection current; magnetic flux density; magnetic resonance electrical impedance tomography; partial differential equation; random noise; reconstruction algorithm; signal-to-noise ratio; spatial resolution; Conductivity; Image reconstruction; Impedance; Magnetic resonance; Noise measurement; Noise reduction; Pollution measurement; Signal to noise ratio; Spatial resolution; Tomography; Denoising; Magnetic Resonance Electrical Impedance Tomography; Partial Differential Equations;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7803-8439-3
Type :
conf
DOI :
10.1109/IEMBS.2004.1403340
Filename :
1403340
Link To Document :
بازگشت