Abstract
This paper presents the principle, structure and performance of a newly developed MR-compatible force sensor. It employs a new optical micrometry that enables highly accurate and highly sensitive displacement measurement. The sensor accuracy is better than 1.0 %, and the maximum displacement of the detector is about 10 μm for a range of the applied force from 0 to 6 N.
Keywords
bio-optics; biomedical MRI; displacement measurement; force sensors; micrometry; optical sensors; MR-compatible optical force sensor; displacement measurement; magnetic resonance imaging; optical micrometry; Displacement measurement; Force sensors; Humans; Image sensors; Lenses; Magnetic resonance imaging; Magnetic sensors; Optical imaging; Optical sensors; Paper technology; MR-compatible force sensor; MRI; Optical micrometry;