Monte Carlo simulation of HiRez PET scanner with Geant4: comparison with experimental data
Author
Rothfuss, H. ; Conti, M. ; Michel, C. ; Panin, V.Y.
Author_Institution
CPS Innovations, Knoxville, TN, USA
Volume
3
fYear
2004
fDate
16-22 Oct. 2004
Firstpage
1867
Abstract
PET scanner simulation is an excellent tool for understanding and making crucial decisions concerning the modeled scanner, such as optimal data organization or rebinning. The simulation is trusted for use only after it has been validated with experimental results. We developed a Geant4 simulation of the CPS HiRez PET scanner. The simulation is benchmarked on experimental results, which are presented in this paper. A point source in the center of the field of view (FOV), a line source in the center of the FOV, and a line source in a scattering phantom have been simulated. Sensitivity, spatial resolution, scatter fraction, and energy spectra have also been studied.
Keywords
Monte Carlo methods; positron emission tomography; CPS HiRez PET scanner; Geant4 simulation; Monte Carlo simulation; PET scanner simulation; data organization; data rebinning; energy spectra; line source; point source; scatter fraction; scattering phantom; sensitivity; spatial resolution; Crystals; Electromagnetic scattering; Geometry; Imaging phantoms; Light scattering; Particle scattering; Positron emission tomography; Solid modeling; Spatial resolution; Technological innovation;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2004 IEEE
Conference_Location
Rome
ISSN
1082-3654
Print_ISBN
0-7803-8700-7
Electronic_ISBN
1082-3654
Type
conf
DOI
10.1109/NSSMIC.2004.1462608
Filename
1462608
Link To Document