• DocumentCode
    438559
  • Title

    Performance evaluation of SIEMENS CTI ECAT EXACT 47 PET scanner USING NEMA NU2-2001

  • Author

    Kim, Jin Su ; Lee, Jae Sung ; Lee, Dong Soo ; Jung, June-Key ; Lee, Myung Chul

  • Author_Institution
    Dept. of Nucl. Med., Seoul Nat. Univ.
  • Volume
    5
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    3118
  • Lastpage
    3120
  • Abstract
    In this study, system performance of Siemens CTI ECAT EXACT 47 PET scanner was measured using NEMA NU2-2001 standard. ECAT EXACT 47 is a BGO crystal based PET scanner. Coincidence window are set to 12 ns, and energy window are set to 250~650 keV. For the spatial resolution measurement, F-18 point source was placed at 1 cm and 10 cm offset from the center of transaxial FOV. Images were reconstructed using FBP with ramp filter. To measure the system sensitivity, NEMA sensitivity phantom filled with F-18 solution and surrounded by 1~5 aluminum sleeves were scanned at the center of transaxial FOV and 10 cm offset from the center. Attenuation free values of sensitivity were estimated by extrapolating data to the zero wall thickness. NEMA scatter phantom with length of 70 cm was filled with F-18 solution (2D: 2,900 MBq), and coincidence count rates were measured for 7 half-lives to obtain noise equivalent count rate (NECR) and scatter fraction. Transverse and axial resolutions at 1 cm offset from the center were 9.1 and 9.8 mm (FBP, 2D and 3D), sensitivity values were 709 (2D), 3332 (3D) cps/MBq at the center and 796 (2D), 4050 (3D) cps/MBq at 10 cm offset from the center. Scatter fractions were 0.19 (2D). Peak true count rate and NECR were 64.0 kcps@37.5 kBq/mL and 49.6 kcps@37.5 kBq/mL in 2D. Information about the performance of EXACT 47 PET scanner reported in this study will be useful for the quantitative analysis of data and determination of optimal image acquisition protocols using this widely used scanner for clinical and research purposes
  • Keywords
    gamma-ray detection; image reconstruction; image resolution; medical image processing; noise; phantoms; positron emission tomography; solid scintillation detectors; 70 cm; BGO crystal; F-18 point source; NEMA NU2-2001 standard; NEMA scatter phantom; Siemens CTI ECAT EXACT 47 PET scanner; aluminum sleeves; axial resolution; clinical purpose; coincidence count rates; coincidence window; energy window; filtered back projection; image reconstruction; noise equivalent count rate; optimal image acquisition protocols; ramp filter; scatter fractions; spatial resolution measurement; transverse resolution; zero wall thickness; Aluminum; Attenuation; Filters; Image reconstruction; Imaging phantoms; Measurement standards; Positron emission tomography; Scattering; Spatial resolution; System performance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • Conference_Location
    Rome
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466341
  • Filename
    1466341