• DocumentCode
    438640
  • Title

    Practical evaluation of an analytical attenuation correction method for a non-uniform attenuator

  • Author

    Ishikawa, Akira ; Ogawa, Koichi

  • Author_Institution
    Dept. of Electon. Informatics, Hosei Univ., Tokyo, Japan
  • Volume
    6
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    3611
  • Abstract
    This paper evaluates an analytical attenuation correction method for a non-uniform attenuator in single photon emission CT. For the analytical attenuation correction method we implemented Kunyansky´s formulation. The performance of this method was compared with that of the OS-EM method with a simulation phantom and an experiment phantom. The results of simulation showed that the analytical method could correct the effect of attenuation at the region where the attenuation coefficients were large. The results of experiments showed that the analytical method was superior to the iterative method and the reduction of scattered photons was very important for quantitative image reconstruction in the clinical setting.
  • Keywords
    attenuators; image reconstruction; iterative methods; phantoms; single photon emission computed tomography; Kunyanskys formulation; OS-EM method; analytical attenuation correction method; experiment phantom; nonuniform attenuator; quantitative image reconstruction; scattered photon; simulation phantom; single photon emission CT; Analytical models; Attenuation; Attenuators; Computed tomography; Electromagnetic scattering; Image analysis; Image reconstruction; Imaging phantoms; Iterative methods; Particle scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466665
  • Filename
    1466665