DocumentCode
438640
Title
Practical evaluation of an analytical attenuation correction method for a non-uniform attenuator
Author
Ishikawa, Akira ; Ogawa, Koichi
Author_Institution
Dept. of Electon. Informatics, Hosei Univ., Tokyo, Japan
Volume
6
fYear
2004
fDate
16-22 Oct. 2004
Firstpage
3611
Abstract
This paper evaluates an analytical attenuation correction method for a non-uniform attenuator in single photon emission CT. For the analytical attenuation correction method we implemented Kunyansky´s formulation. The performance of this method was compared with that of the OS-EM method with a simulation phantom and an experiment phantom. The results of simulation showed that the analytical method could correct the effect of attenuation at the region where the attenuation coefficients were large. The results of experiments showed that the analytical method was superior to the iterative method and the reduction of scattered photons was very important for quantitative image reconstruction in the clinical setting.
Keywords
attenuators; image reconstruction; iterative methods; phantoms; single photon emission computed tomography; Kunyanskys formulation; OS-EM method; analytical attenuation correction method; experiment phantom; nonuniform attenuator; quantitative image reconstruction; scattered photon; simulation phantom; single photon emission CT; Analytical models; Attenuation; Attenuators; Computed tomography; Electromagnetic scattering; Image analysis; Image reconstruction; Imaging phantoms; Iterative methods; Particle scattering;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2004 IEEE
ISSN
1082-3654
Print_ISBN
0-7803-8700-7
Electronic_ISBN
1082-3654
Type
conf
DOI
10.1109/NSSMIC.2004.1466665
Filename
1466665
Link To Document