A mixed continuous-pulsed reset technique for digitized preamplifiers of radiation signals
Author
Zocca, F. ; Pullia, A. ; Boiano, C. ; Bassini, R.
Author_Institution
Dept. of Phys., Milan Univ., Italy
Volume
1
fYear
2005
fDate
23-29 Oct. 2005
Firstpage
382
Lastpage
386
Abstract
We have realized a mixed continuous-pulsed reset preamplifier suitable for multi-channel gamma-ray sensors. The continuous-reset mode is used for typical signal amplitudes within the ADC input range, or 1 V. The pulsed-reset mode works on larger signal amplitudes, either caused by individual events or by bursts of piled-up events. Such operation mode maximizes the live time in the typical energy range and minimizes the dead time caused by ADC saturation on large individual signals or at high count rates. The amplitude of large individual signals can also be estimated from the reset time still achieving a high resolution. Using this technique and a bulky high purity germanium detector we got a "continuous-reset" range of 2 MeV with a resolution of 2.02 keV fwhm at 1.408 MeV, and an extended "pulsed-reset" range of 27 MeV, with a typical resolution of 0.3% at 27 MeV.
Keywords
gamma-ray apparatus; germanium radiation detectors; nuclear electronics; preamplifiers; 1.408 MeV; 2 MeV; 2.02 keV; 27 MeV; ADC saturation dead time; analog-to-digital converter; continuous-reset mode; digitized preamplifiers; high purity germanium detector; mixed continuous-pulsed reset preamplifier; multichannel gamma-ray sensors; pulsed-reset mode; Amplitude estimation; Capacitance; Detectors; Dynamic range; Energy resolution; Germanium; Nuclear physics; Nuclear power generation; Preamplifiers; Signal resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2005 IEEE
ISSN
1095-7863
Print_ISBN
0-7803-9221-3
Type
conf
DOI
10.1109/NSSMIC.2005.1596275
Filename
1596275
Link To Document