Title :
Radial position sensing in a coplanar-grid high-pressure xenon gamma-ray spectrometer
Author :
Kiff, Scott D. ; He, Zhong ; Tepper, Gary C.
Author_Institution :
Michigan Univ., Ann Arbor, MI
Abstract :
This paper describes a technique for determining the interaction radius of a gamma ray within a cylindrical detector. This can be useful information for correcting pulse amplitudes and selecting events of interest. A theoretical expression for the event radius is derived based on anode signals, then simulations are compared with experimental results
Keywords :
gamma-ray detection; gamma-ray spectrometers; gas scintillation detectors; position sensitive particle detectors; anode signals; coplanar-grid high-pressure xenon gamma-ray spectrometer; cylindrical detector; event radius; interaction radius; pulse amplitude correction; radial position sensing; Anodes; Cathodes; Gamma ray detection; Gamma ray detectors; Pulse measurements; Solid scintillation detectors; Spectroscopy; Strips; Time measurement; Xenon;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2005 IEEE
Conference_Location :
Fajardo
Print_ISBN :
0-7803-9221-3
DOI :
10.1109/NSSMIC.2005.1596464