Title :
Exact ROI image reconstruction with perturbed source trajectories in C-arm CT
Author :
Xia, Dan ; Sidky, Emil Y. ; Yu, Lifeng ; Zou, Yu ; Pan, Xiaochuan
Author_Institution :
Dept. of Radiol., Chicago Univ., IL, USA
Abstract :
In C-arm CT, the actual trajectory executed by the X-ray source may deviate from the designed scan due to mechanical error. Without taking into account this deviation, artifacts will appear in the reconstructed image. In this work, we propose to use the recently developed backprojection filtration (BPF) algorithm, which can accommodate general trajectories, to incorporate the deviation of the source trajectory. The preliminary studies demonstrate that this algorithm provides the exact image reconstruction in the case that the deviation of the source trajectory from the ideal trajectory can be measured.
Keywords :
computerised tomography; image reconstruction; medical image processing; C-arm CT; X-ray source; backprojection filtration algorithm; exact ROI image reconstruction; perturbed source trajectories; source trajectory deviation; Algorithm design and analysis; Band pass filters; Biomedical applications of radiation; Computed tomography; Image converters; Image reconstruction; Prototypes; Reconstruction algorithms; Surgical instruments; Vibrations;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2005 IEEE
Print_ISBN :
0-7803-9221-3
DOI :
10.1109/NSSMIC.2005.1596712