A High Precision Peak Detect Sample and Hold Circuit
Author
Chang, P.Y. ; Chou, H.P.
Author_Institution
Dept. of Eng. & Syst. Sci., National Tsing Hua Univ., Hsinchu
Volume
1
fYear
2006
fDate
Oct. 29 2006-Nov. 1 2006
Firstpage
329
Lastpage
331
Abstract
The paper presents a high precision peak detect sample and hold (PDSH) circuit for nuclear radiation spectroscopy. A small compensation current source is applied to reduce the measuring error and improve the resolution. For nuclear pulses with the peak value ranging from 0.1 to 1 volts with a frequency about 100 kHz, test results showed that voltage drop is 2.16muV/mus and the voltage error is less than 0.5 mV.
Keywords
nuclear electronics; peak detectors; sample and hold circuits; high precision peak detect sample and hold circuit; nuclear pulses; nuclear radiation spectroscopy; CMOS technology; Capacitors; Circuit testing; Diodes; Mirrors; Modems; Radiation detectors; Spectroscopy; Temperature dependence; Voltage;