• DocumentCode
    462517
  • Title

    A New Dual Anode Microstrip Plate Chamber for X-rays: Experimental Results

  • Author

    Carita, Liliana P M ; do Carmo, S.J.C. ; Santos, Filomena P. ; Conde, C.A.N.

  • Author_Institution
    Departamento de Fisica, Univ. de Coimbra
  • Volume
    3
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    1349
  • Lastpage
    1353
  • Abstract
    A novel design for a gaseous radiation detector is described. In uses a new microstructure which has an anode surrounded by another close-lying second anode, playing the role of a grid at a potential between anode and cathode that allows a more accurate definition of the multiplication volume. The main characteristics of this new detector are described and its performance discussed. The best energy resolutions for 5.9 keV X-rays obtained were 19.5% for Xe and 14.3% for P10 mixture.
  • Keywords
    X-ray detection; proportional counters; 5.9 keV; P10 gas mixture; X-ray detection; dual anode microstrip plate chamber; gaseous radiation detector; multiplication volume; xenon gas; Anodes; Cathodes; Electrons; Energy resolution; Fluctuations; Geometry; Microstrip; Microstructure; Radiation detectors; X-rays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.354153
  • Filename
    4179266