Abstract :
We suggest a simple and efficient approach to reduce, and, in some cases, eliminate the windmill artifact. Our method does not require hardware changes. The idea is to improve the z-sampling by applying up-sampling in the detector row direction. One recent result show that the linear interpolation can be improved by shifting the samples by some small amount; we utilize this approach for our purposes. Evaluation shows that using some specified amount of the shift in interpolation, the proposed method provides a significant improvement of the windmill artifact with the clinical data, without a noticeable loss of z-resolution.
Keywords :
computerised tomography; image sampling; interpolation; medical image processing; computed tomography; linear interpolation; up-sampling; windmill artifact elimination; windmill correction; z-sampling; Biomedical imaging; Computed tomography; Data processing; Detectors; Equations; Hardware; Interpolation; Kernel; Nuclear and plasma sciences; Silver;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location :
San Diego, CA
ISSN :
1095-7863
Print_ISBN :
1-4244-0560-2
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2006.354371
Filename :
4179485
Link To Document :
بازگشت