DocumentCode :
468622
Title :
A novel approach to design of uUniplanar gradient coil for fully open MRI system
Author :
Li, Xia ; Xie, Dexin ; Wang, Jinming ; Bai, Baodong
Author_Institution :
Shenyang Univ. of Technol., Shenyang
fYear :
2007
fDate :
8-11 Oct. 2007
Firstpage :
1867
Lastpage :
1870
Abstract :
A new approach to the design of uniplanar gradient coils for fully open magnetic resonance imaging (MRI) system is proposed to build a specified gradient magnetic field in a target region. In this new methodology, the surface current density of the coil is represented by a two-dimensional Fourier series expansion. A cost function is constructed in terms of the summation of squared error and a penalty function. Minimizing the cost function, the continuous current density distribution for the uniplanar gradient coil is obtained. Applying the stream function technique to the resulting continuous current distribution, discrete current patterns can be generated. Employing the Biot-Savart law to the discrete current loops, the gradient magnetic field has been re-evaluated in order to validate the method proposed. Using this approach, a uniplanar x-gradient coil with high level of linearity is designed.
Keywords :
Fourier series; coils; current density; magnetic fields; magnetic resonance imaging; cost function; current density distribution; fully open MRI system; gradient magnetic field; magnetic resonance imaging; stream function technique; surface current density; two-dimensional Fourier series expansion; uniplanar gradient coil; Coils; Cost function; Current density; Current distribution; Fourier series; Fourier transforms; Integral equations; Linearity; Magnetic fields; Magnetic resonance imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electrical Machines and Systems, 2007. ICEMS. International Conference on
Conference_Location :
Seoul
Print_ISBN :
978-89-86510-07-2
Electronic_ISBN :
978-89-86510-07-2
Type :
conf
Filename :
4412113