Title :
Calibration of pinhole microSPECT system using line sources
Author :
Zhou, Rong ; Ma, Tianyu ; Liu, Yaqiang ; Sun, Xishan ; Kang, Xiaowen ; Wu, Zhaoxia ; Wang, Shi ; Jing, Yongjie ; Cheng, Jianping
Author_Institution :
Tsinghua Univ., Beijing
fDate :
Oct. 26 2007-Nov. 3 2007
Abstract :
A method using line sources to calibrate a pinhole MicroSPECT system was proposed. Projections\´ coordinates of "vitual" point sources were obtained by solving equations of centerlines, which were detected out of projection images of line sources by Radon transform. Effect of depth of interaction that gamma photons rip into crystals was modeled when calculating the theoretical position of projection. Several systems were simulated by Monte Carlo method and calibrated both by the proposed method and by using ideal point sources for comparison. Calibration accuracy achieved by the proposed method is comparable to that by using ideal point sources.
Keywords :
Monte Carlo methods; Radon transforms; calibration; single photon emission computed tomography; Monte Carlo method; Radon transform; calibration; line sources; pinhole MicroSPECT; Calibration; Detectors; Equations; Geometry; Nuclear and plasma sciences; Photonic crystals; Physics; Sun; Telephony; Transforms;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
Conference_Location :
Honolulu, HI
Print_ISBN :
978-1-4244-0922-8
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2007.4436853