DocumentCode :
469847
Title :
Development of a high resolution whole-body DOI PET system
Author :
Ohtani, Atsushi ; Tanaka, Kazumi ; Kitamura, Keishi ; Mizuta, Tetsuro ; Inoue, Yoshihiro ; Tonami, Hiromichi ; Ohi, Junichi
Author_Institution :
Shimadzu Corp., Kyoto
Volume :
5
fYear :
2007
fDate :
Oct. 26 2007-Nov. 3 2007
Firstpage :
3695
Lastpage :
3699
Abstract :
High spatial resolution Positron Emission Tomography (PET) imaging provides functional data about a patient´s body and provides us with a range of new possibilities, including medicine development, new diagnostic methods, and others. The use of Depth of Interaction (DOI) detectors is one effective technique for the improvement of whole-body PET imaging, as is evident in the fact that most small animal PET images are taken using a DOI system. In this study, we developed a prototype whole-body DOI- equipped PET system and evaluated its performance. This scanner´s DOI detectors consist of dual-layer (front and rear) 9 x 10 array of GSO/GSO crystals, a light guide and two rectangular double-anode photo multiplier tubes (PMTs). The DOI layers are discriminated according to decay time differences, which are controlled by Ce concentrations. The DOI detectors are arranged in a circular detector ring, with a diameter of 664 mm. In order to avoid parallax errors, coincidental events, including rear layer detection, were addressed to the nearest neighbor LOR front-layer coincidence pair. To evaluate the spatial resolution of this PET system at various radial positions, 22Na point sources at 1, 10, 20, and 25 cm offset from the center were reconstructed, both with and without DOI information. Results showed that spatial resolution degradation at 25 cm improved from 195 % to 154 % as a result of using DOI data. During a visual evaluation using a Derenzo phantom, image reconstruction using DOI data clarified the image and corrected hot rod shapes. The results presented here show that DOI detectors were useful in creating an effective whole-body PET system.
Keywords :
image reconstruction; image resolution; medical image processing; phantoms; photomultipliers; positron emission tomography; solid scintillation detectors; Derenzo phantom; Gd2SiO5; circular detector ring; depth of interaction detectors; double-anode photomultiplier tubes; high resolution whole-body DOI PET system; image reconstruction; light guide; nearest neighbor LOR front-layer coincidence pair; positron emission tomography; spatial resolution; Animals; Biomedical imaging; Detectors; High-resolution imaging; Image reconstruction; Medical diagnostic imaging; Positron emission tomography; Prototypes; Spatial resolution; Whole-body PET;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
Conference_Location :
Honolulu, HI
ISSN :
1095-7863
Print_ISBN :
978-1-4244-0922-8
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2007.4436924
Filename :
4436924
Link To Document :
بازگشت