Abstract :
We characterized a two-photon method to calibrate photon-counting detectors. We verified this method by comparison to a national primary standard detector scale. This comparison showed agreement of the two methods to 0.14(14) % (k=1).
Keywords :
calibration; photodetectors; photon counting; two-photon processes; correlated-photon calibration technique; independent verification; photon-counting detector calibration; two-photon method; Calibration; Delay; Detectors; Educational institutions; Electronic equipment testing; NIST; Photonics; Power measurement; Pulse measurements; Uncertainty;