DocumentCode
472252
Title
Two-Channel Bioimpedance Monitor for Impedance Cardiography
Author
Vondra, Vlastimil ; Halamek, Josef ; Viscor, Ivo ; Jurak, Pavel
Author_Institution
Inst. of Sci. Instrum., Acad. of Sci., Brno
fYear
2006
fDate
Aug. 30 2006-Sept. 3 2006
Firstpage
6061
Lastpage
6063
Abstract
In this paper we introduce our conception of a two-channel bioimpedance monitor for the impedance cardiography. We describe the design of a homemade device which is based on a direct digital synthesizer and a digital down converter. This solution enables us to obtain a high-quality bioimpedance signal, which can be used for computing the cardiac output beat per beat. It further enables us to analyze impedance phase changes beat per beat. Our device allows simultaneous measuring of complex impedance in two arbitrary directions on human/animal body without any interference between channels. The setting of the main parameters of the measurement like the amplitude and the frequency of the measurement current depends on the user
Keywords
bioelectric potentials; biomedical equipment; biomedical measurement; direct digital synthesis; electrocardiography; DDS; bioimpedance measurement; digital down converter; direct digital synthesizer; homemade device design; impedance cardiography; two-channel bioimpedance monitor; Animals; Bioimpedance; Cardiography; Current measurement; Frequency measurement; Humans; Impedance measurement; Interference; Monitoring; Synthesizers;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
Conference_Location
New York, NY
ISSN
1557-170X
Print_ISBN
1-4244-0032-5
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2006.260011
Filename
4463190
Link To Document