Title :
The Changes in Pulse Transit Time at Specific Cuff Pressures during Inflation and Deflation
Author :
Liu, Yinbo ; Poon, Carmen C Y ; Zhang, Y.T.
Author_Institution :
Dept. of Electron. Eng., Chinese Univ. of Hong Kong, Shatin
fYear :
2006
fDate :
Aug. 30 2006-Sept. 3 2006
Firstpage :
6404
Lastpage :
6405
Abstract :
The changes in pulse transit time (PTT) during the continuous slow deflation of brachial cuff were early reported; however, the PTTs obtained for specific cuff pressures during inflation or deflation have not been compared before. Therefore, the objective of this study is to examine the differences in PTT when cuff pressure (Peun) was raised or deflated to the desired level. Sixteen subjects participated in this study and according to their systolic blood pressure (SBP) and diastolic blood pressure (DBF), 8 levels of Pcuff were predetermined for them individually. Pcuff was directly raised to each predetermined level while 20 seconds of electrocardiographic and photoplethysmographic signals were recorded for the calculation of PTT. Another set of recordings were taken when Pcuff was raised above the SBP and deflated to the predetermined levels. The results of this study showed that PTT increase significantly when Pcuff was larger than 80% of DBF, regardless of whether Pcuff was reached by inflation or deflation. Overall, no significant difference was found between PTT obtained during inflation and deflation for 12 out of the 16 subjects. To conclude, changes in PTT are mainly induced by the level of cuff pressure when there is no prolonged period of artery occlusion
Keywords :
blood vessels; electrocardiography; plethysmography; pneumodynamics; 20 s; SBP; brachial cuff; deflation; diastolic blood pressure; electrocardiographic signal; inflation; photoplethysmographic signal; pulse transit time; systolic blood pressure; Aging; Arteries; Blood pressure; Brachytherapy; Cities and towns; Electrocardiography; Force measurement; Pulse measurements; USA Councils; Ultrasonic variables measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
Conference_Location :
New York, NY
ISSN :
1557-170X
Print_ISBN :
1-4244-0032-5
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2006.260410
Filename :
4463276
Link To Document :
بازگشت