DocumentCode :
473748
Title :
Reliable automated QT interval measurement for clinical evaluation
Author :
Chen, X. ; Xie, W. ; Dong, L. ; Tin, ZK ; Sankaranarayanan, M.
Author_Institution :
Inst. for Infocomm Res., Nanyang
fYear :
2006
fDate :
17-20 Sept. 2006
Firstpage :
373
Lastpage :
376
Abstract :
The aim of this study was to develop a reliable automated method of QT interval measurement for ECG signals, and to participate the PhysioNet/Computers in Cardiology Challenge 2006. A scheme based on Kalman filter was proposed to conduct beat segmentation and searching of the first representative beat from ECG waveform, and a wavelet based algorithm was used to detect the onset of QRS complex and the offset of T wave which yield the measurement ofQT interval. Our preliminary QT measurement results have been submitted to CinC for performance evaluation and further improvement to the algorithms and system is carried out.
Keywords :
Kalman filters; electrocardiography; medical computing; performance evaluation; ECG signals; ECG waveform; Kalman filter; QRS complex; T wave; automated QT interval measurement; beat segmentation; clinical evaluation; first representative beat; performance evaluation; wavelet based algorithm; Automatic testing; Cardiology; Costs; Electrocardiography; Gaussian noise; Heart rate interval; Noise measurement; Pollution measurement; Time measurement; Tin;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology, 2006
Conference_Location :
Valencia
Print_ISBN :
978-1-4244-2532-7
Type :
conf
Filename :
4511866
Link To Document :
بازگشت