• DocumentCode
    47610
  • Title

    Research on High Accuracy Current Comparator and Self-Calibration Methods

  • Author

    He Xiaobing ; Wang Wei ; Zhang Xin ; Dai Dongxue

  • Author_Institution
    Nat. Inst. of Metrol., Beijing, China
  • Volume
    62
  • Issue
    6
  • fYear
    2013
  • fDate
    Jun-13
  • Firstpage
    1669
  • Lastpage
    1674
  • Abstract
    This paper introduces the structure and the working principle of a high-accuracy current comparator. This paper also indicates the primary source of error: the leakage capacitance coupling current between the turns of the secondary winding. In order to reduce this error, the secondary winding is wound with a coaxial cable. With good shielding, the error introduced by the connecting windings is minimized. This paper also presents a new self-calibration method for the 1:1 current comparator and an innovative design of the current comparator. The self-calibration method can also be used to calibrate multiratio current comparators with an uncertainty of less than 5×10-7 at a primary current ranging from 5 A to 2000 A.
  • Keywords
    calibration; electric current measurement; current 5 A to 2000 A; high accuracy current comparator; leakage capacitance coupling current; primary source of error; secondary winding; self-calibration method; Current comparator; leakage capacitance; self-calibration;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2013.2253978
  • Filename