DocumentCode :
482438
Title :
Small valley gap magnet design for a 10MeV PET cyclotron
Author :
Qin, Bin ; Yang, Jun ; Liu, Kaifeng ; Huang, Zhenling ; Zhao, Liangchao ; Chu, Chengjie ; Fan, Mingwu
Author_Institution :
Huazhong Univ. of Sci. & Technol., Wuhan
fYear :
2008
fDate :
17-20 Oct. 2008
Firstpage :
906
Lastpage :
909
Abstract :
Low energy compact cyclotrons applied in Positron Emission Tomography (PET) have foreseeable prospects with the growing demands in medical applications. Compared with conventional dasiadeep valleypsila magnet type, the cyclotron magnet with small valley gap (SVG) has higher average magnetic field which reduces magnet pole size, but smaller field flutter also decreases the axial focusing of beam. Considering that the main purpose of PET cyclotrons is to produce short-lived isotopes which need moderate beam intensity and quality, the SVG type magnet can fulfill the requirements. And beam loss due to Lorentz strip caused by high magnetic field is insignificant with low energy beam. This paper describes the design and considerations of SVG type magnet including the central region for a 10 MeV PET cyclotron. Mechanical considerations and field mapping system are introduced.
Keywords :
cyclotrons; positron emission tomography; Lorentz strip; PET cyclotron; SVG type magnet; axial focusing; beam loss; cyclotron magnet; deep valley magnet type; electron volt energy 10 MeV; field flutter; field mapping system; low energy compact cyclotrons; magnet pole size; magnetic field; positron emission tomography; small valley gap magnet design; Biomedical imaging; Cyclotrons; Hydrogen; Ion sources; Isotopes; Magnetic fields; Medical diagnostic imaging; Positron emission tomography; Production; Radio frequency;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electrical Machines and Systems, 2008. ICEMS 2008. International Conference on
Conference_Location :
Wuhan
Print_ISBN :
978-1-4244-3826-6
Electronic_ISBN :
978-7-5062-9221-4
Type :
conf
Filename :
4770842
Link To Document :
بازگشت