DocumentCode :
482439
Title :
The mechanical instability of the BPMs at HLS
Author :
Cao, Y. ; Sun, B.G. ; Li, J.W. ; He, D.H. ; Wang, X.H. ; Lu, P. ; Li, C. ; Xuan, K.
Author_Institution :
NSRL, Univ. of Sci. & Technol. of China, Hefei
fYear :
2008
fDate :
17-20 Oct. 2008
Firstpage :
910
Lastpage :
913
Abstract :
Beam position monitor (BPM) used to measure the position of the electron beam in storage ring. The instability of the BPM itself will definitely influence the beam position readout. At Hefei Light Source (HLS), the mechanical instability of the BPM was studied with raster gauges. By tracing the displacement of the BPM chamber, it was observed that the BPM chamber moved from time to time. The movement corresponded to its surface temperature variation, which was caused by the synchrotron light. This paper recorded this phenomenon and gave the proper explanation.
Keywords :
electron beams; mechanical stability; particle beam diagnostics; synchrotron radiation; BPM; HLS; Hefei light source; beam position monitor; electron beam; mechanical instability; raster gauges; storage ring; surface temperature variation; synchrotron light; Cooling; Electron beams; Feedback; High level synthesis; Magnets; Stability; Storage rings; Sun; Synchrotrons; Temperature sensors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electrical Machines and Systems, 2008. ICEMS 2008. International Conference on
Conference_Location :
Wuhan
Print_ISBN :
978-1-4244-3826-6
Electronic_ISBN :
978-7-5062-9221-4
Type :
conf
Filename :
4770843
Link To Document :
بازگشت