Title :
An Ultrasonic-Adaptive Beamforming Method and Its Application for Trans-skull Imaging of Certain Types of Head Injuries; Part I: Transmission Mode
Author :
Shapoori, Kiyanoosh ; Sadler, Jeff ; Wydra, Adrian ; Malyarenko, Eugene V. ; Sinclair, Anthony N. ; Maev, Roman Gr
Author_Institution :
Dept. of Mech. & Ind. Eng., Univ. of Toronto, Toronto, ON, Canada
Abstract :
A new adaptive beamforming algorithm for imaging via small-aperture 1-D ultrasonic-phased arrays through composite layered structures is reported. Such structures cause acoustic phase aberration and wave refraction at undulating interfaces and can lead to significant distortion of an ultrasonic field pattern produced by conventional beamforming techniques. This distortion takes the form of defocusing the ultrasonic field transmitted through the barrier and causes loss of resolution and overall degradation of image quality. To compensate for the phase aberration and the refractional effects, we developed and examined an adaptive beamforming algorithm for small-aperture linear-phased arrays. After accurately assessing the barrier´s local geometry and sound speed, the method calculates a new timing scheme to refocus the distorted beam at its original location. As a tentative application, implementation of this method for trans-skull imaging of certain types of head injuries through human skull is discussed. Simulation and laboratory results of applying the method on skull-mimicking phantoms are presented. Correction of up to 2.5 cm focal point displacement at up to 10 cm depth under our skull phantom is demonstrated. Quantitative assessment of the method in a variety of temporal focusing scenarios is also reported. Overall temporal deviation on the order of a few nanoseconds was observed between the simulated and experimental results. The single-point adaptive focusing results demonstrate strong potential of our approach for diagnostic imaging through intact human skull. The algorithms were implemented on an ultrasound advanced open-platform controlling 64 active elements on a 128-element phased array.
Keywords :
array signal processing; biomedical ultrasonics; injuries; medical image processing; phantoms; acoustic phase aberration; diagnostic imaging; head injuries; human skull; single-point adaptive focusing; skull-mimicking phantoms; small-aperture 1D ultrasonic-phased arrays; trans-skull imaging; ultrasonic-adaptive beamforming method; wave refraction; Acoustic distortion; Acoustics; Array signal processing; Focusing; Phased arrays; Timing; Adaptive beamforming; Focusing; Phase aberration; Refraction; Trans-skull Imaging; ULA-OP; Ultrasonic phased array; focusing; phase aberration; refraction; trans-skull imaging; ultrasonic phased array; ultrasound advanced open-platform (ULA-OP);
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :