    Spectral stability of 0.9 µm standard radiation thermometer

  • Author

    Sakuma, Fumihiro

  • Author_Institution
    NMIJ, Nat. Inst. of Adv. Ind. Sci. & Technol., Tsukuba, Japan
  • fYear
    2009
  • fDate
    18-21 Aug. 2009
  • Firstpage
    2797
  • Lastpage
    2801
  • Abstract
    Long term spectral characteristics of 0.9 mum radiation thermometers were studied by using a monochromator and also fixed point blackbodies. Some thermometers showed shifts in their center wavelengths. Normally the thermometers were stable within plusmn 0.2 nm per year. The trends of the two measurements were the same but they had a small difference in the absolute values. The spectral stability of the radiation thermometers can be monitored by the signal ratio between two fixed point blackbodies.
  • Keywords
    blackbody radiation; monochromators; spectral methods of temperature measurement; thermometers; fixed point blackbodies; long-term spectral characteristics; monochromator; signal ratio; spectral stability; standard radiation thermometer; wavelength 0.9 mum; Aluminum; Calibration; Copper; Instruments; Interference; Optical filters; Temperature; Thermal stability; Uncertainty; Wavelength measurement; fixed-point blackbodies; interference filter; long-term stability; radiation thermometer; spectral responsivity;