DocumentCode
50701
Title
Development of a Measurement Setup for High Impulse Currents
Author
Istrate, D. ; Blanc, Isabelle ; Fortune, Dave
Author_Institution
Lab. Nat. de Metrol. et d´essai (LNE), Trappes, France
Volume
62
Issue
6
fYear
2013
fDate
Jun-13
Firstpage
1473
Lastpage
1478
Abstract
The understanding of high-current events such as short-circuit transients and impulse currents is important for power quality management. The work carried out and presented in this paper is focused on the characterization of commercial current transducers for traceable on-site measurements of 8/20 μs (rise/fall time) impulse currents up to 50 kA peak value with a target uncertainty of 0.1% for the complete measurement system. The investigation method consists of using two current transducers based on different technologies that will detect simultaneously the same impulse current.
Keywords
electric current measurement; measurement uncertainty; power supply quality; transducers; current transducer; impulse current measurement; measurement setup development; measurement system; measurement uncertainty; power quality management; traceable on-site measurement; Calibration; Current measurement; Current transformers; Frequency response; Sensors; Transducers; Uncertainty; Comparison method; Pearson current transformer; Rogowski coil; impulse current;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2013.2239018
Filename
6459016
Link To Document