DocumentCode :
53731
Title :
Uncertainty Quantification in Transcranial Magnetic Stimulation via High-Dimensional Model Representation
Author :
Gomez, Luis J. ; Yucel, Abdulkadir C. ; Hernandez-Garcia, Luis ; Taylor, Stephan F. ; Michielssen, Eric
Author_Institution :
Dept. of Electr. Eng. & Comput. Sci., Univ. of Michigan, Ann Arbor, MI, USA
Volume :
62
Issue :
1
fYear :
2015
fDate :
Jan. 2015
Firstpage :
361
Lastpage :
372
Abstract :
A computational framework for uncertainty quantification in transcranial magnetic stimulation (TMS) is presented. The framework leverages high-dimensional model representations (HDMRs), which approximate observables (i.e., quantities of interest such as electric (E) fields induced inside targeted cortical regions) via series of iteratively constructed component functions involving only the most significant random variables (i.e., parameters that characterize the uncertainty in a TMS setup such as the position and orientation of TMS coils, as well as the size, shape, and conductivity of the head tissue). The component functions of HDMR expansions are approximated via a multielement probabilistic collocation (ME-PC) method. While approximating each component function, a quasi-static finite-difference simulator is used to compute observables at integration/collocation points dictated by the ME-PC method. The proposed framework requires far fewer simulations than traditional Monte Carlo methods for providing highly accurate statistical information (e.g., the mean and standard deviation) about the observables. The efficiency and accuracy of the proposed framework are demonstrated via its application to the statistical characterization of E-fields generated by TMS inside cortical regions of an MRI-derived realistic head model. Numerical results show that while uncertainties in tissue conductivities have negligible effects on TMS operation, variations in coil position/orientation and brain size significantly affect the induced E-fields. Our numerical results have several implications for the use of TMS during depression therapy: 1) uncertainty in the coil position and orientation may reduce the response rates of patients; 2) practitioners should favor targets on the crest of a gyrus to obtain maximal stimulation; and 3) an increasing scalp-to-cortex distance reduces the magnitude of E-fields on the surface and inside the cortex.
Keywords :
biomedical MRI; medical disorders; medical signal processing; uncertainty handling; ME-PC method; MRI derived realistic head model; TMS coils; depression therapy; head tissue; high dimensional model representations; iteratively constructed component functions; multielement probabilistic collocation; transcranial magnetic stimulation; uncertainty quantification; Brain modeling; Coils; Computational modeling; Conductivity; Random variables; Uncertainty; Generalized polynomial chaos (gPC); high-dimensional model representation (HDMR); magnetic stimulation; multielement probabilistic collocation (ME-PC); sensitivity analysis; surrogate model; transcranial magnetic stimulation (TMS); uncertainty quantification;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2014.2353993