Title :
Design and Research of a Corneal Grafting Robotic System
Author :
Yang, Yang ; Shijing, Deng ; Jingjing, Xiao ; Cunliang, Xu
Author_Institution :
Beihang Univ., Beijing, China
Abstract :
Due to the demands for accuracy and stability in corneal grafting microsurgery, it is rather difficult for surgeon to deal with it independently. To achieve corneal grafting automation, the robotic system for corneal grafting microsurgery is proposed and designed preliminarily using modularization. The robot prototype and the end-effectors (trephine and gripper) are developed, and the overall schematic diagram of the system is put forward. Finally, the experiments of trephination, suturing and knot-tying manipulation are implemented, the last two are based on the planning methods for suturing theoretical trajectory and knot-tying sequences. Meantime, the Stitching force-testing data is also obtained. The results show that the proposed robotic system is feasible in the automation of corneal grafting microsurgery.
Keywords :
end effectors; eye; grippers; medical robotics; prosthetics; surgery; corneal grafting microsurgery; corneal grafting robotic system; end effector; knot tying manipulation; modularization; robot prototype; stitching force testing data; suturing theoretical trajectory; trephination; corneal grafting; knot-tying manipulation; microsurgical robot; suture; trephination;
Conference_Titel :
Digital Manufacturing and Automation (ICDMA), 2010 International Conference on
Conference_Location :
ChangSha
Print_ISBN :
978-0-7695-4286-7
DOI :
10.1109/ICDMA.2010.29