DocumentCode
539800
Title
Performance Test of Steel Tube Enclosed Thermistors in Air Temperature Measurements
Author
Chen Hua ; Peng Ji ; Yan Feng ; Sui Yongxin ; Yang Huaijiang
Author_Institution
State Key Lab. of Appl. Opt., Chinese Acad. of Sci., Changchun, China
Volume
2
fYear
2011
fDate
6-7 Jan. 2011
Firstpage
854
Lastpage
857
Abstract
The performances of two steel tube enclosed precision thermistors are tested in a well controlled air environment to determine if the thermistors have enough accuracy for ±0.01°C levels of air temperature gradient measurement. Three Fluke´s 5641 and three 5610 thermistors were tested for 24 hours in Fluke´s 7008 calibration bath in the first step. In the second step, two of the 5641s were airproofed in a temperature well controlled tube, the performances of these two thermistors in steady state and transient state air temperature environments were tested for 8 hours separately. The results illustrate that, when steel tube enclosed precision thermistors are used in air temperature measurement, there are unstable departures from each other with a maximum 8.8 m°C in constant air medium and 27 m°C in temperature changing air medium. The transient-time errors are too large to show the true amounts of temperature changes for air temperature measurement.
Keywords
calibration; measurement errors; temperature measurement; thermistors; air temperature gradient measurement; calibration; steel tube enclosed precision thermistor; transient-time error; Electron tubes; Probes; Steel; Temperature measurement; Temperature sensors; Thermal stability; Thermistors; Air Medium; Steel Tube Enclosed; Temperature Measurement; Thermistor;
fLanguage
English
Publisher
ieee
Conference_Titel
Measuring Technology and Mechatronics Automation (ICMTMA), 2011 Third International Conference on
Conference_Location
Shangshai
Print_ISBN
978-1-4244-9010-3
Type
conf
DOI
10.1109/ICMTMA.2011.496