• DocumentCode
    541563
  • Title

    The evaluation of methods in determination of the arterial compliance for real-time application

  • Author

    Hu, Weichih ; Shyu, Liang-Yu ; Cheng, Hou-Ming ; Chen, Chen-Huan

  • Author_Institution
    Chung Yuan Christian Univ., Chungli, Taiwan
  • fYear
    2010
  • fDate
    26-29 Sept. 2010
  • Firstpage
    373
  • Lastpage
    376
  • Abstract
    The objective of this study is to evaluate a method that could be used in real-time arterial compliance estimation while measure blood pressure noninvasively. The noninvasive pressure waveform at carotid artery and pressure volume relation (PVR) at brachial artery were first evaluated to assert the consistency of compliance calculation. The arterial pressure waveform was acquired using oscillometric method. The result was checked against the data calculated using invasive pressure waveforms at aortic root and brachial artery that were acquired simultaneously. The result showed that the relaxation constant could be used to reveal stiffness of artery. The index could be assessed using pulse waveform acquired oscillometric method.
  • Keywords
    biomechanics; blood pressure measurement; blood vessels; elasticity; arterial compliance determination; brachial artery; carotid artery; noninvasive blood pressure measurement; noninvasive pressure waveform; oscillometric method; pressure volume relation; real time arterial compliance estimation; Arteries; Blood pressure; Brachytherapy; Correlation; Indexes; Integrated circuit modeling; Pressure measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computing in Cardiology, 2010
  • Conference_Location
    Belfast
  • ISSN
    0276-6547
  • Print_ISBN
    978-1-4244-7318-2
  • Electronic_ISBN
    0276-6547
  • Type

    conf

  • Filename
    5737987