Automated QT interval measurement in Holter ECGs recorded at 180 and 1000 samples/second
Author
Panicker, G.K. ; Salvi, V. ; Karnad, D.R. ; Macfarlane, P. ; Clark, E. ; Ramasamy, A. ; Kothari, S. ; Narula, D.
Author_Institution
Quintiles Cardiac Safety Services, Mumbai, India
fYear
2010
fDate
26-29 Sept. 2010
Firstpage
761
Lastpage
764
Abstract
To study the effect of sampling rate on automated QT measurements, Holter ECGs were recorded at 180 and 1000 samples/second (s/s) using 2 recorders; 30 ECG snapshots were extracted at varying heart rates from 16 healthy subjects and re-sampled to 180, 500 or 1000s/s using the Antares software. QT interval by CalECG algorithm was longer (5.0±6.3 ms, p<;0.001) in 180s/s ECGs than in 1000s/s ECGs. It decreased to 2.1±5.8ms when 180s/s ECGs were re-sampled to 500s/s, and to 2.6±6.2ms at 1000s/s. It also decreased progressively on resampling both sets of ECGs to 1000s/s (2.6±6.2ms), 500s/s (1.8±5.5ms) and 180s/s (0.4±5.9ms). Differences in QT interval were independent of the QT measurement algorithm used: University of Glasgow (Uni-G) program and CalECG for 500s/s ECGs; Veritas and CalECG for 1000s/s ECGs. Thus, QT interval is longer in ECGs with lower sampling rates; resampling them to a higher resolution partially compensates for this.
Keywords
biomedical measurement; electrocardiography; software packages; Antares software; CalECG algorithm; ECG snapshots; Holter ECG; automated QT interval measurement; Drugs; Electrocardiography; Gold; Heart rate; Software; Software algorithms; Software measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Computing in Cardiology, 2010
Conference_Location
Belfast
ISSN
0276-6547
Print_ISBN
978-1-4244-7318-2
Type
conf
Filename
5738084
Link To Document