Title :
Optimization of the size sample in the elastic moduli measurement of human cortical bone
Author :
Darmana, R. ; Hobatho, M.C. ; Ashman, RB ; Morucci, J.P.
Author_Institution :
INSERM, Centre Hospitalier Hotel Dieu, Toulouse, France
fDate :
Oct. 29 1992-Nov. 1 1992
Abstract :
A mechanical method using strain gages has been developed to measure the elastic moduli of human cortical bone of the tibia. The average of elastic moduli measured were found to be for the radial modulus : E1 = 9.5 ± 0.4 GPa, for the tangential modulus : E2 = 11.5 ± 1.1 GPa and for the axial modulus: E3 = 21.3 ± 1.2 GPa. The technique has been validated by comparing the results with that of the ultrasonic measurements performed on samples of another human tibia. T-test shows no significant differences between the results.
Keywords :
biomechanics; biomedical measurement; biomedical ultrasonics; bone; elastic moduli; elastic moduli measurement; optimisation; T-test; axial modulus; elastic moduli measurement; human cortical bone; optimization; radial modulus; size sample; strain gage; tangential modulus; tibia; ultrasonic measurement; Bones; Displacement measurement; Humans; Size measurement;
Conference_Titel :
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location :
Paris
Print_ISBN :
0-7803-0785-2
Electronic_ISBN :
0-7803-0816-6
DOI :
10.1109/IEMBS.1992.5760835