• DocumentCode
    544268
  • Title

    Non-contact measurement of thermal properties of the skin

  • Author

    Togawa, Tatsuo ; Saito, Hirokazu

  • Author_Institution
    Inst. for Med. & Dental Eng., Tokyo Med. & Dental Univ., Tokyo, Japan
  • Volume
    1
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    232
  • Lastpage
    233
  • Abstract
    The square root of product of thermal conductivity, density and specific heat, V(kpc), of the skin was estimated from infrared radiation measurement at stepwise change of ambient radiation temperature. To effect this change, two shades at different temperatures were switched mechanically. Measurements were made at different sites of the body in 20 subjects, and averaged V(kpc) was 1884 ± 611(SD) W s1/2 / m2 K. Application of this method to thermographic measurement is discussed.
  • Keywords
    biomedical measurement; biothermics; density; infrared imaging; skin; specific heat; ambient radiation temperature; density; infrared radiation measurement; noncontact measurement; skin thermal property; specific heat; thermal conductivity; thermographic measurement; Forehead; Heating; Radiometry;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5760941
  • Filename
    5760941