DocumentCode :
544381
Title :
An innovative two-channel rate and morphology method for complex arrhythmia diagnosis in impiantatile devices
Author :
Chiang, Chih-ming James ; Jenkins, Janice M. ; DiCarlo, Lorenzo A.
Author_Institution :
Deft of EECS, University of Michigan, Ann Arbor, Michigan, 48109-2122, USA
Volume :
2
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
496
Lastpage :
497
Abstract :
Impiantatile cardioverter-defìbrillators (ICDs) are highly effective in preventing sudden cardiac death. However, false-positive diagnoses cause inappropriate device shocks estimated to be as high as 40%. To address this problem, the Augmented Two-Channel Arrhythmia Detection [A2CAD] algorithm, a real-time scheme utilizing information from both the atrial and ventricular channels, is introduced. The algorithm uses rate detection as a first stage and augments this with morphological signal analysis in rhythms that confound the rate-only diagnoses. Results from 40 patient cases indicate 100% success rate for A2CAD, validating its potential for implementation in future implantable devices.
Keywords :
Morphology;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location :
Paris, France
Print_ISBN :
0-7803-0785-2
Electronic_ISBN :
0-7803-0816-6
Type :
conf
DOI :
10.1109/IEMBS.1992.5761077
Filename :
5761077
Link To Document :
بازگشت