DocumentCode :
544465
Title :
Accurate calibration of pneumotacographs using a syringe and polynomial curve fitting
Author :
Neto, Antonio Giannella ; Vale, Marcio J O do ; Melo, Marcos F.Vidal
Author_Institution :
Biomedicai Engineering Program, COPPE, Federal University of Rio de Janeiro, PO Box 68510, 21945 Rio de Janeiro, RJ, Brazil
Volume :
2
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
693
Lastpage :
694
Abstract :
This work presents a simple, reliable and accurate calibration technique for pneumotacographs. It is based on a polynomial curve fitting which minimizes the calculated volume error for a set of injections of a calibrated syringe. In conditions similar to human spirometry the results meet the standards of the American Thoracic Society.
Keywords :
Converters; Fitting;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location :
Paris, France
Print_ISBN :
0-7803-0785-2
Electronic_ISBN :
0-7803-0816-6
Type :
conf
DOI :
10.1109/IEMBS.1992.5761174
Filename :
5761174
Link To Document :
بازگشت