• DocumentCode
    544483
  • Title

    Changes in the finger artery compliance and elasticity as a function of temperature

  • Author

    Zhang, Sijian ; Patterson, Robert ; Cohn, Jay N. ; Wang, Li

  • Author_Institution
    University of Minncsota, MN 55455, U.S.A.
  • Volume
    2
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    730
  • Lastpage
    731
  • Abstract
    An improved noninvasive electrical impedance finger cuff was designed to measure the finger artery compliance (Ca = ΔV/ΔP) and the volume elastic modulus (Ev = Ya/Ca), where AV — pulsatile volume change; ΔV = pulse pressure; Va-mean arterial volume. Voluntary cardio-respiratory synchronization (VCRS) and stepwise cuff pressure release techniques were used to obtain the finger artery compliances after hand immersion at temperatures of 20°C5 30°C and 40°C, As a result, the finger artery compliance Ca increased as the temperature increased, whereas its elasticity Ev remained relatively constant.
  • Keywords
    Atmospheric measurements; Cardiology; Particle measurements;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761192