Evaluation of the dynamic response of catheter-manometer systems
Author :
Tuohy, B. ; Sheahan, N.F. ; Kirkham, R. ; Coakley, D. ; Malone, J.P.
Author_Institution :
Dept of Medical Physics and Bioengineering, St. James´´s Hospital, Dublin 8
Volume :
6
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
2400
Lastpage :
2401
Abstract :
The "fast flush test" is commonly used to determine the dynamic response of catheter-manometer systems in-vivo. This test was compared to an established bench top test In vitro. Estimates for the resonant frequency and damping factor of the measurement systems were made for each test. There was a significant linear correlation between the results of the two tests. The fast flush test provides a useful estimate of the dynamic properties of catheter-manometer systems.