• DocumentCode
    544690
  • Title

    Video endoscopy: Our experience with image quality assessment

  • Author

    McMahon, B.P. ; Hegarty, F. ; Malone, J.F.

  • Author_Institution
    Dept. of Med. Phys. & Bio-Eng., St. James Hosp., Dublin, Ireland
  • Volume
    3
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    1171
  • Lastpage
    1172
  • Abstract
    With over two years experience in the support of a busy endoscopy unit the authors have gathered important information which brings into question the supplier recommendations for set up and calibration of video endoscopic systems . Errors in colour rendition and resolution due to electronic drift, subjective set up, user errors and variation between different imaging systems suggest the need for quality assurance procedures and the development of a set of standards.
  • Keywords
    calibration; endoscopes; quality assurance; calibration; colour rendition errors; electronic drift; image quality assessment; quality assurance procedures; user errors; video endoscopy; Biology; Switches;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761436
  • Filename
    5761436